      Specifications:

        Wxx02 (2000 N (450lbs) Version) Wxx04 (4000 N (900 lbs) Version)
      Voltage
      Input voltage [VDC] 12 24 12 24
      Load
      Static load (Fx), maximum [N (lbs)] 2000 (450) 4000 (900)
      Dynamic load (Fx), maximum [N (lbs)] 2000 (450) 4000 (900)
      Stroke
      Stroke lengths, standard [cm (inch)] 10 (3.94), 20 (7.87), 30 (11.81), 40 (15.75), 50 (19.69)
      Speed
      Speed version 58A 54A 58A 54A 58A 58A
      Speed, no load [mm/s (inch/s)] 5.8 (0.23) 11.0 (0.43) 5.8 (0.23) 11.0 (0.43) 5.8 (0.23) 5.8 (0.23)
      Speed, rated load [mm/s (inch/s)] 4.0 (0.16) 8.0 (0.31) 4.0 (0.16) 8.0 (0.31) 4.0 (0.16) 4.0 (0.16)
      Current
      Current draw @ rated max. load [A] 4.5 9.0 2.2 4.5 9.0 4.5
      Current draw @ stall/in-rush [A] 14.0 21.0 8.0 11.0 21.0 11.0
      General Data
      End play, maximum [mm (inch)] 0.5 (0.020)
      Operating temperature limits [°C (°F)] -25 to +40 (-13 to +104) (1)
      Duty cycle, maximum [%] 10
      On-time, maximum [s] 180 90 180 90 90 90
      Life, average [cycles] 10,000
      Sound level [dBa] < 45
      Lead screw type acme
      Protection class IP67
      Anti rotation mechanism(2) yes
      Analog feedback output signal voltage [VDC] 0.5 - 4.5
      Analog feedback output signal linearity [%] 0.5
      Digital feedback input voltage [VDC] 3.8 - 24
      Digital feedback output resolution [pulses/mm] 10 5 10 5 10 10
      Compliances IEC 60601-1, CSA C22.2, AAMI ES60601-1 (3)
      Options
      • Electronic limit switches (ELS) (4)
      • ELS + low level switching (1) (5)
      • Analog or digital position feedback, with or without ELS
      • Mounting adapters turned 90°

      ?           1) Thrust is reduced at lower temperatures for 2000 N (450 lbs) high speed (54A) and 4000 N (900 lbs) units with ELS + low level switching (option E).
                 2) Prevents the extension tube from rotating if it is not fixed in the end.
                 3) 4 kN units greater than 30 cm of stroke are not ETL recognized. ETL certification available on N, E, P and D feedback options only.
                 4) Shuts off power at the end of stroke and all along the stroke at overload conditions. ELS is normally set for 120% of the rated dynamic maximum load.
                 5) Low level switching allows low level signals to control the actuator movement instead of having to switch the high amperage motor voltage.